Alat Uji Lingkungan Jawa Barat - Depok

  • 02177827903
  • 081284403633

BENCHTOP WATER QUALITY METER


  • Harga: CALL

  • Minimal Pembelian: 1 unit

  • Pengiriman: 3 hari sejak pembayaran

  • Lokasi: Jawa Barat - Depok

  • Dilihat: 2.730 Kali

  • Pembaruan: 18-04-2024 16:24

Benchtop Water Quality Meter
Type : PC-80
Product : Italy
Fungsi : Alat uji kualitas air, pH, mV, Conductivity, TDS, Salinity, Resitivity, Temperature

Spesifikasi :
pH measuring range : -2...20
Resolution/ Relative accuracy : 0.1 / 0.01 / 0.001 / ±0,002
Points of calibration : 1...5
Buffer recognized : USA, NIST, DIN, 5 USER
Indication of calibration points : YES
GLP Calibration : YES with data and time
CAL DUE with allarm : YES
Criteria of measurement stability : Low-Medium-High
mV measuring range/ Resolution : ± 2000 / 0.1 / 1
Conductivity measuring range : 0,00...20,00 - 200,0 - 2000 μS /
  : 2,00 - 20,00 - 200,0 - 1000 mS
Resolution : Automatic scale
Relative accuracy : ±1 % full scale
TDS measuring range : 0,1mg/L...500 gr/L
TDS Factor / Relative accuracy : 0.40...1.00 /±1 % full scale
Salinity range/ Relative accuracy : 0,01...100ppt /±1 % full scale
Resistivity range / Relative accuracy : 0...20MΩ /±1 % full scale
Temperature range ºC / Accuracy : -20,0...120,0 / ± 0,2°C
Temperature compensation : Manual or automatic (NTC 30KΩ) 0-100°C
Temperature calibration : YES
Calibration points COND : 1...4
Calibration standard recognized : 84, 1413 μS / 12.88, 111.8 mS 
    1 point user defined
Indication of calibration standard : YES
TR Reference temperature : 15...30ºC
GLP Calibration : YES with data and time
Indication of cell condition : YES with icon
CAL DUE COND with allarm : YES
Type of cell : 2 or 4 pole
Cell constant : 0.1...10.0
System GLP : YES
Memory : MAN/AUTO 1000 date with data and time
Display : Dot matrix multicolors with icons, 
  : digital and analogic
Inputs : 2 x BNC, 2 x Cinch/RCA (ATC), 
    USB for printer or PC,
    USB for external keyboard
Language : 8 languages
Manual On-line : YES
Magnetical stirrer : YES
Speed controll stirrer : 0...3000 rpm
IP Protection : IP 54
Power supply : AC/DC adapter 9 V / 300 mA
Dimensions / Weight : 360 x 220 x 100 mm / 1250 g
Complete unit : pH electrode Standard S7
  : Cable S7/BNC
  : Conductivity cell glass VPT 80/1
  : Colered pH buffer solution pH7 & 4
  : Standard conductivity solution 1413µS and 12,88 mS
  : Power supply with stirrer and Carrying Case.